Research Papers: Friction & Wear

A Novel Elastic Squeeze Film Total Hip Replacement

[+] Author and Article Information
S. Boedo

Department of Mechanical Engineering,
Rochester Institute of Technology,
Rochester, NY 14623

J. F. Booker

Sibley School of Mechanical
and Aerospace Engineering,
Cornell University,
Ithaca, NY 14853

Contributed by the Tribology Division of ASME for publication in the Journal of Tribology. Manuscript received September 15, 2012; final manuscript received June 25, 2013; published online August 6, 2013. Assoc. Editor: Mihai Arghir.

J. Tribol 136(1), 011101 (Aug 16, 2013) (12 pages) Paper No: TRIB-12-1150; doi: 10.1115/1.4024968 History: Received September 15, 2012; Revised June 25, 2013

This paper describes a new approach to the design of total hip replacements with the goal of enhancing lubrication and reducing wear. Elastic elements and ellipsoidal cup surface geometry are incorporated into the new design to promote and enhance ‘squeeze-film’ action over ‘wedge-film’ action employed in conventional artificial hip joints. Employing an established finite element lubrication model with a realistic gait cycle and realistic ball-to-cup clearance specifications, it is found that significantly larger minimum film thicknesses and significantly smaller maximum film pressures are predicted over the stance-phase portion of the gait cycle when compared with conventional designs.

Copyright © 2014 by ASME
Your Session has timed out. Please sign back in to continue.

References

Ingham, E., and Fisher, J., 2005, “The Role of Macrophages in Osteolysis of Total Joint Replacement,” Biomaterials, 26, pp. 1271–1286. [CrossRef] [PubMed]
Firkins, P. J., Tipper, J. L., Saadatzadeh, M. R., Ingham, E., Stone, M. H., Farrar, R., and Fisher, J., 2001, “Quantitative Analysis of Wear and Wear Debris From Metal-on-Metal Hip Prostheses Tested in a Physiological Hip Joint Simulator,” Bio-Med. Mater. Eng., 11, pp. 143–157.
Catelas, I., Medley, J. B., Campbell, P. A., Huk, O. L., and Bobyn, J. D., 2004, “Comparison of In Vitro With In Vivo Characteristics of Wear Particles From Metal-on-Metal Hip Implants,” J. Biomed. Mater. Res B: Appl. Biomater., 70, pp. 167–178. [CrossRef] [PubMed]
Brown, C., Fisher, J., and Ingham, E., 2006, “Biological Effects of Clinically Relevant Wear Particles From Metal-on-Metal Hip Prostheses,” Proc. IMechE H: J. Eng. Med., 220, pp. 355–369. [CrossRef]
Cobb, A. G., and Schmalzreid, T. P., 2006, “The Clinical Significance of Metal Ion Release From Cobalt-Chromium Metal-on-Metal Hip Prostheses,” Proc. IMechE H: J. Eng. Med., 220, pp. 385–398. [CrossRef]
Bhamra, M. S., and Case, C. P., 2006, “Biological Effects of Metal-on-Metal Hip Replacements,” Proc. IMechE H: J. Eng. Med., 220, pp. 379–384. [CrossRef]
Feder, B. J., 2008, “That Must Be Bob. I Hear his New Hip Squeaking,” The New York Times, Business section, May 11, 2008.
Goenka, P. K., 1980, “Effect of Surface Ellipticity on Dynamically Loaded Spherical and Cylindrical Joints and Bearings,” Ph.D. thesis, Cornell University, Ithaca, NY.
Kothari, M., Booker, J. F., and Bartel, D. L., 1995, “Analysis of Artificial Joints as Spherical Bearings,” Lubricants and Lubrication, D.Dowson, C.Taylor, T.Childs, G.Dalmaz, eds., Elsevier, pp. 93–98.
Jin, Z. M., and Dowson, D., 1999, “A Full Numerical Analysis of Hydrodynamic Lubrication in Artificial Hip Joint Replacements Constructed From Hard Materials,” Proc. IMechE C: J. Mech. Eng. Sci., 213, pp. 355–370. [CrossRef]
Wang, F. C., and Jin, Z. M., 2007, “Effect of Non-Spherical Bearing Geometry on Transient Elastohydrodynamic Lubrication in Metal-on-Metal Hip Joint Implants,” Proc. IMechE J: J. Eng. Tribol., 221, pp. 379–389. [CrossRef]
Wang, F. C., and Jin, Z. M., 2008, “Transient Elastohydrodynamic Lubrication of Hip Joint Implants,” ASME J. Tribol., 130, p. 011007. [CrossRef]
Wang, F. C., Zhao, S. X., Felix Quinonez, A., Xu, H., Mei, X. S., and Jin, Z. M., 2009, “Nonsphericity of Bearing Geometry and Lubrication in Hip Joint Replacements,” ASME J. Tribol., 131, p. 031201. [CrossRef]
Meng, Q. E., Liu, F., Fisher, J., and Jin, Z. M., 2011, “Transient Elastohydrodynamic Lubrication Analysis of a Novel Metal-on-Metal Hip Prosthesis With a Non-Spherical Femoral Bearing Surface,” Proc. IMechE H: J. Eng. Med., 225, pp. 25–37. [CrossRef]
Dowson, D., and Jin, Z. M., 2006, “Metal-on-Metal Hip Joint Tribology,” Proc. IMechE J: J. Eng. Tribol., 220, pp. 107–118. [CrossRef]
Fan, J., Myant, C. W., Underwood, R., Cann, P. M., and Hart, A., 2011, “Inlet Protein Aggregation: A New Mechanism for Lubricating Film Formation With Model Synovial Fluids,” Proc. IMechE H: J. Eng. Med., 225, pp. 696–709. [CrossRef]
Sprecher, C., Hauert, R., Grad, S., Tager, G., Fisher, A., and Shneider, E., 2003, “Solid Lubrication-A Relevant Lubrication Mechanism for Reducing Wear in Metal-on-Metal THA Components?,” 49th Annual Meeting, ORS, p. 1391.
Unsworth, A., 1975, “Cavitation in Human Joints,” Cavitation and Related Phenomena in Lubrication, D.Dowson, M.Godet, C. M.Taylor, eds., IMechE Publications Ltd., pp. 119–127.
Kumar, A., and Booker, J. F., 1991, “A Finite Element Cavitation Algorithm,” ASME J. Tribol., 113, pp. 276–286. [CrossRef]
Boedo, S., and Booker, J. F., 2001, “Finite Element Analysis of Elastic Engine Bearing Lubrication: Application,” Revue Européenne des Éléments Finis, 10, pp. 725–740.
Booker, J. F., and Boedo, S., 2001, “Finite Element Analysis of Elastic Engine Bearing Lubrication: Theory,” Revue Européenne des Éléments Finis, 10, pp. 705–724.
International Standards Organization, 2002, ISO 14242-1: Implants for Surgery—Wear of Total Hip-Joint Prostheses—Part 1: Loading and Displacement Parameters for Wear Testing Machines and Corresponding Environmental Conditions for Test.
Cooke, A. V., Dowson, D., and Wright, V., 1978, “The Rheology of Synovial Fluid and Some Potential Synthetic Lubricants for Degenerate Synovial Joints,” Eng. Med., 7, pp. 66–72. [CrossRef]
Yao, J. Q., Laurent, M. P., Johnson, T. S., Blanchard, C. R., and Crowinshield, R. D., 2003, “The Influence of Lubricant and Material on Polymer/CoCr Sliding Friction,” Wear, 255, pp. 780–784. [CrossRef]
Mattei, L., Di Puccio, F., Piccigallo, B., and Ciulli, E., 2011, “Lubrication and Wear Modelling of Artificial Hip Joints: A Review,” Tribol. Int., 44, pp. 532–549. [CrossRef]
Booker, J. F., and Huebner, K. H., 1972, “Application of Finite Element Methods to Lubrication: An Engineering Approach,” ASME J. Lub. Tech., 94, pp. 313–323. [CrossRef]
Goenka, P. K., and Booker, J. F., 1980, “Spherical Bearings: Static and Dynamic Analysis via the Finite Element Method,” ASME J. Lub. Tech., 102, pp. 308–318. [CrossRef]

Figures

Grahic Jump Location
Fig. 1

Sample embodiment of squeeze film artificial hip joint

Grahic Jump Location
Fig. 2

Surface geometry of rigid cup portion

Grahic Jump Location
Fig. 3

ISO 14242 duty cycle

Grahic Jump Location
Fig. 4

Spatial view of lubricant film mesh

Grahic Jump Location
Fig. 5

Projected view of lubricant film mesh

Grahic Jump Location
Fig. 6

Effect of ellipticity on time histories of minimum film thickness and maximum film pressure

Grahic Jump Location
Fig. 7

Pressure distributions at t = 0.5 s: R1 = 16 mm, μ = 2.5 mPa-s

Grahic Jump Location
Fig. 8

Effect of ellipticity on bearing performance: R1 = 16 μm, μ = 2.5 mPa-s

Grahic Jump Location
Fig. 9

Effect of ellipticity on bearing performance: R1 = 25 μm, μ = 2.5 mPa-s

Grahic Jump Location
Fig. 10

Effect of initial conditions: R1 = 16 mm, μ = 2.5 mPa-s, C0 = 30 μm, δ = 40 μm

Grahic Jump Location
Fig. 11

Film mesh: cup design with wide slot excluded

Grahic Jump Location
Fig. 12

Effect of wide slot on bearing performance

Grahic Jump Location
Fig. 13

Bearing performance comparison: R1 = 14 mm, C0 = 30 μm

Grahic Jump Location
Fig. 14

Effect of ellipticity on periodic time histories of minimum film thickness and maximum film pressure: complete hemispherical cup

Grahic Jump Location
Fig. 15

Planar fluid film element

Grahic Jump Location
Fig. 16

Hemispherical bearing under axisymmetric pure squeeze

Grahic Jump Location
Fig. 17

Mesh refinement study: hemispherical bearing under pure squeeze

Grahic Jump Location
Fig. 19

Column diameter required for elastic load of 350 N at eZ= 0

Tables

Errata

Discussions

Some tools below are only available to our subscribers or users with an online account.

Related Content

Customize your page view by dragging and repositioning the boxes below.

Related Journal Articles
Related eBook Content
Topic Collections

Sorry! You do not have access to this content. For assistance or to subscribe, please contact us:

  • TELEPHONE: 1-800-843-2763 (Toll-free in the USA)
  • EMAIL: asmedigitalcollection@asme.org
Sign In